GROUPA 2 LYS
These reagents kits are in vitro diagnostic medical devices (IVDMD) for professional use. They are used for ABO-RH1 grouping and/or RH-K phenotyping of human red blood cells. They are not intended for screening for weak antigen D.
These devices, intended for ABO grouping, must be used with HEMALYS or red blood cells magnetized according to the recommendations of DIAGAST for the realization of the reverse typing test (reverse group). These devices are for use on the automated machines QWALYS® in the E.M.® Technology method.
Groupa 2 Lys enable determination of the presence of erythrocytic antigens A (ABO1), B (ABO2) and D (RH1).
Groupa 2 Lys (DIAGAST Ref.: 79964) is a kit consisting of:
- 12 Groupa 2 Lys microplates for 12 x 8 ABO-RH1 grouping.
Each microplate contains 8 wells of anti-A, anti-B, anti-A,B, anti-D1, anti-D2 and negative control, and 8 x 6 empty wells. - 1 x 40-mL vial of MagneLys.
NB: The clones used for anti-D1 and anti-D2 give negative results in case of DVI. For this specific detection (DVI), use Groupa DVI Lys.
Groupa 2 Lys | 79964 | Technical insert |
MagneLys | 79900 | Technical insert |
BROMELINE 5X100ML | 69860 | Technical insert |
HEMALYS 1 A1, A2, B, O | 59661 | Technical insert |
D-Plates | 79851 | Technical insert |